An experience of the clinical study with angiography system using a Flat Panel Detector
Kenji Okusako, Masachika Shogaki, Kouji Yokoyama, Takayoshi Ogawa, Katsuhisa Tanaka, Takao Ichida, Kenji Kishimoto, Masakatsu Hatagawa, Kazuo Okuyama, Hiroaki Kudoh, Saori Tanaka, Kenji Nakamura, Sigeyuki Ikeda, Katsumi Suzuki
Author information
JOURNAL FREE ACCESS

2002 Volume 22 Issue 4 Pages 255-263

Details
Abstract
We had the chance of clinical studies by new proto-type Digital Subtraction Angiography (DSA) System with the Flat Panel Detector (FPD).
That system has two types of detector, FPD and I. I. -CCD camera, and we can compare with image quality in clinical.
We measured and discussed about the physical characteristic data of both detectors. Up to physical characteristic evaluation, FPD has good MTF and SNR performance, comparing with I.I.-CCD camera. In clinical evaluation, DSA image of FPD was superior to that of I. I. -CCD camera by visual comparison of medical doctors. Finally we discussed about the feasibility of replacement of detectors. We expect that DSA system will accept FPD in near future.