Relationship between line spread function (LSF), or slice sensitivity profile (SSP), and poin t spread function (PSF) in CT image system
Masaki OhkuboShinichi WadaTeiji KobayashiLee YongbumTsai Du-Yih
Author information
JOURNAL FREE ACCESS

2004 Volume 24 Issue 3 Pages 115-122

Details
Abstract

In the CT image system, we revealed the relationship between line spread function (LSF), or slice sensitivity profile (SSP), and point spread function (PSF). In the system, the following equation has been reported; I(x, y) = 0(x, y)** PSF(x, y), in which I(x, y) and 0(x, y) are CT image and object function, respectively, and ** is 2-dimensional convolution. In the same way, the following 3-dimensional expression applies; I' (x, y, z) = 0' (x, y, z) * * * PSF' (x, y, z), in which z-axis is the direction perpendicular to the x/y-scan plane. We defined that the CT image system was separable, when the above two equations could be transformed into following equations; I(x, y) = [0(x, y)*LSF(x)]* LSFy(y) and I' (x, y, z) = [0' (x, y, z) * SSP(z)]** PSF(x, y), respectively, in which LSF(x) and LSF, (y) are LSFs in x- and y-direction, respectively. Previous reports for the LSF and SSP are considered to assume the separable-system. Under the condition of separable-system, we derived following equations; PSF (x, y) = LSF, (x)LSFy(y) and PSF' (x, y, z) = PSF(x, y) SSP(z). They were validated by the computersimulations. When the study based on 1-dimensional functions of LSF and SSP are expanded to that based on 2- or 3-dimensional functions of PSF, derived equations must be required.

Content from these authors
© The Japan Society of Medical Physics
Previous article
feedback
Top