Japanese Journal of Magnetic Resonance in Medicine
Online ISSN : 2434-0499
Print ISSN : 0914-9457
TECHNICAL NOTE
A Phantom Study to Investigate Fat Suppression in DWIBS
Hiroaki MINAMIMitsuyuki TAKAHASHIHirofumi HATAYoshito NAKAJIMA
Author information
JOURNAL FREE ACCESS

2020 Volume 40 Issue 3 Pages 118-127

Details
Abstract

 Diffusion-weighted whole body imaging with background body signal suppression uses the short time inversion (TI) recovery (STIR) method, which is effective against the inhomogeneity of a magnetic field, for fat suppression. Oil causes chemical shifts with different transfer amounts, owing to the change in its chemical structure, and the null points of CH2 and olefins present different values. Consequently, fat cannot be completely suppressed by the STIR method, and it is necessary to combine it with other fat suppression methods. In this study, using phantom imaging proposed by the authors with SSRF, both 1.5T and 3.0T fat suppression effects were improved. The null point was prolonged using CHESS with STIR. In this case, the fat suppression effect decreased with the same TI as that of the STIR.

Content from these authors
© 2020 Japanese Society for Magnetic Resonance in Medicine
Previous article
feedback
Top