Japanese Journal of Magnetic Resonance in Medicine
Online ISSN : 2434-0499
Print ISSN : 0914-9457
Scientific Exhibit Award of The 49th Annual Meeting
Implementation of Three-Dimensional Fast Spin Echo Sequences on the MRI Simulator [Presidential Award Proceedings]
Katsumi KOSERyoichi KOSE
Author information
JOURNAL OPEN ACCESS

2022 Volume 42 Issue 2 Pages 44-49

Details
Abstract

 As the three-dimensional fast spin echo (3D FSE) imaging uses greater than 100 refocusing pulses which generate several higher-order echoes, its Bloch simulation is challenging. The simulation was performed using a numerical phantom with 256 cube voxels and 320 subvoxels (isochromats) in the readout direction. The simulation was performed using approximately 1.58 billion isochromats and the calculation time was nearly 15 hours using an RTX 2080Ti graphics processing unit having 11 GB memory. The calculated result demonstrated that the 3D FSE was correctly simulated.

Content from these authors
© 2022 Japanese Society for Magnetic Resonance in Medicine

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja
Previous article Next article
feedback
Top