Japanese Journal of Magnetic Resonance in Medicine
Online ISSN : 2434-0499
Print ISSN : 0914-9457
NOTE
Non-contrast MR Angiography of the Renal Artery with Golden Angle Radial Sparse Parallel-Volumetric Interpolated Breath-hold Examination : Optimal Imaging Conditions
Kazuyuki ITONozomi HAMASAKIKazuhiko DORYOYutaka OZAKI
Author information
JOURNAL OPEN ACCESS

2023 Volume 43 Issue 4 Pages 174-181

Details
Abstract

 Golden Angle Radial Sparse Parallel-Volumetric Interpolated Breath hold Examination (GRASP-VIBE) using compressed sensing and golden-angle radial sampling is a gradient echo (GRE) sequence that provides an inflow effect. The purpose of this study was to investigate parameters for non-contrast visualization of the renal artery using GRASP-VIBE with respiratory synchronization by the liver gate. We used a flow phantom and a self-made phantom to determine effective fat suppression methods and appropriate settings for repetition time (TR) and flip angle (FA). We also performed imaging on five healthy volunteers and compared the ability to visualize the renal artery. In the phantom study, water excitation showed the highest fat suppression effect ; the flow phantom was well visualized at TR 7.2 ms~9.0 ms and FA 9°~13°. In the volunteer study, the renal artery was best visualized at TR 7.2 ms and FA 13°. The results suggest that GRASP-VIBE can be used to visualize the renal artery without contrast enhancement by using the parameter settings examined in this study.

Content from these authors
© 2023 Japanese Society for Magnetic Resonance in Medicine

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja
Previous article Next article
feedback
Top