Japanese Journal of Oral and Maxillofacial Surgery
Online ISSN : 2186-1579
Print ISSN : 0021-5163
ISSN-L : 0021-5163
Invited review article
Mandible reconstruction using a patient-compatible intracorporeal fixation plate (TruMatch Reconstruction system®)
Michihiro UEDAShin RINTakashi NIIYAMA
Author information
JOURNAL FREE ACCESS

2024 Volume 70 Issue 8 Pages 320-326

Details
Abstract
 After continuous disarticulation of the mandible, reconstruction with a free osteocutaneous flap with vascular is effective. However, problems such as occlusal deviation 16.7% (4/24) and facial deformity 8.3% (2/24) may occur in the medium- to long-term course. In particular, when the muscle process and mandibular nucleus are resected and the proximal residual bone is small, dislocation is likely to occur, and intraoperative positioning of the jaw is not only difficult, but also tends to change over time.
 The TruMatch Reconstruction system is a new system that has become available in Japan due to the development of surgical simulation software and the progress of 3D printers. This system is not only accurate in mandibular reconstruction, but also enables accurate resection of tumours according to the preoperative plan by using a cutting guide. It also reduces operative time and blood loss, and is considered to be an effective item for mandibular reconstruction.
 However, there are issues that need to be improved in the future, such as the low degree of freedom of the plate and the long time between planning and completion of the product.
Content from these authors
© 2024 Japanese Society of Oral and Mxillofacial Surgeons
Previous article Next article
feedback
Top