Physicochemical Characteristics of Antimony Microelectrode with Special Reference to Selection of Standard Buffers
Yutaka MATSUMURANoriko SATAKEMamoru FUJIMOTO
Author information
JOURNAL FREE ACCESS

1980 Volume 30 Issue 4 Pages 509-528

Details
Abstract
1) Antimony (Sb) microelectrodes with tips of 2 to 5 μm in outside diameter were constructed, and their electromotive forces (EMF) were tested in response to the pH of several calibration buffers. The data were compared with those of glass pH electrodes.
2) Increasing ionic strength (I) caused a significant deviation in pH readings of the Sb-microelectrode (ΔpHSb-Glass). The above salt effect was empirically given by
ΔpHSb-Glass=0.017-0.125I (I=0.09-0.22)
3) Even with the correction for ionic strength, the values of ΔpHSb-Glass were still dependent on the nature of calibration buffers employed. Among various buffers, Tris buffer [Tris (hydroxymethyl)-aminomethane, (TRAM)] had the least disturbance on the Sb-microelectrode, and bicarbonate had also relatively minor effect.
4) The pH value according to the Sb-electrode was caused to deviate considerably by phosphate.The effect of phosphate within the range of 1 to 67 mM can be predicted by
1/ΔpHSb-Glass=1+ (0.174/[PO4])
5) Like phosphate, BES [N, N-Bis (2-hydroxyethyl)-2-aminoethane sulfonic acid] also had a significant effect on the EMF of the Sb-microelectrode.
6) Upon consideration of the above facts, measurements of several biological fluids, such as those of the in vivo pH of intratubular fluid of bullfrog kidney were carried out with the Sb-microelectrode calibrated by 90 mM Tris buffer.
7) Thus, it was concluded that the Sb-microelectrode when properly calibrated could be used for the pH measurement of ultraminute amounts of biological samples.