JOURNAL OF THE JAPANESE ASSOCIATION OF RURAL MEDICINE
Online ISSN : 1349-7421
Print ISSN : 0468-2513
ISSN-L : 0468-2513
RESEARCH REPORT
Evaluation of Usefulness of X-ray Tube Protector for X-ray Fluoroscopy in ERCP etc
Ryuuta YOSHIDAKiyoshi KITAJIMAKatsuyuki CHICHIDATakayuki HONDANoritaka KAWAKAMIKouichi ISHIMORIHiroyuki SUDOUKensei KAZAMAHidenori MASHIKOMiyuki ADUMATakaaki MATSUDA
Author information
JOURNAL FREE ACCESS

2014 Volume 63 Issue 2 Pages 127-131

Details
Abstract
  The purpose of the present study was to reduce absorbed dose rates in air by means of the X-ray tube protector in the X-ray fluoroscopy room. We measured absorbed dose rates in air using an ionization chamber type survey meter, made the scatter radiation maps and worked out the reduction rate of absorbed dose rates in air. Absorbed dose rate in air reduced 75-90% with use of the X-ray tube protector. Especially, absorbed dose rate in air reduced 85-90% at the height of crystalline lens and thyroid gland of the X-ray technician near the patient. It was suggested X-ray protector could reduce the risk of radiation exposure of radiologists effectively.
Content from these authors
© 2014 THE JAPANESE ASSOCIATION OF RURAL MEDICINE