STUDY ON THE UNLINEARITY OF THE DOSE RATE WITH THE TUBE CURRENT OF THE DEEP THERAPY APPARATUS
KATUHIKO YAMADAMASAJIRO TAKIUCHIIKUO KIMURA
Author information
JOURNAL FREE ACCESS

1959 Volume 15 Issue 2 Pages 147-150

Details
Abstract
When we measure the out put dose (air dose) of the deep therapy apparatus at the fixed peak voltage, we experience the unlinearity of the dose rate with the increasing tube current. We thought that this unlinearity was caused by the following two reasons. (1) The measured dose rate dose not increase expectedly, because the ion recombination in the ionization chamber becomes high with the increasing dose rate. (2) The ripples of the tube voltage increase with the increasing tube current. Then the effective value of the tube voltage decreases, following the decreasing dose rate. From our experiments we decided that the origin of the above mentioned unlinearity was this decreasing effective voltage caused by the ripples.