X-RAY CINE MAGNIFICATION ANGIOGRAPHY BY 0.6mm FOCAL SPOT
ISAO TANAKATAKASHI WAKAMATSUTOSHIYA SANO
Author information
JOURNAL FREE ACCESS

1983 Volume 39 Issue 2 Pages 148-153

Details
Abstract
To observe microstructures on the X-ray cinematography by 0.6mm focal spot, usefulness of enlarged shooting was studied. As a result, even at a focus of 0.6mm, the enlarged shooting (ca. 1.7-fold) revealed the superior total M.T.F. than that of contact photography. Furthermore, Groedel effect upon magnification was as good as the grid system. In the cardiac catheterization by 0.6mm focal spot being employed most frequently at present, the enlarged shooting without using grid is a useful method. Thus, it is considered to take up this method as much as possible.