X-ray Conditions and Response Characteristics of Automatic Dose Control in Cinematography
HIROAKI ARAI
Author information
JOURNAL FREE ACCESS

1997 Volume 53 Issue 11 Pages 1673-1682

Details
Abstract
X-ray characteristics including subject thickness(copper plate), tube voltage, tube current and irradiation time were measured at stability, with an automatic dose control x-ray generator for cineangiography. Regardless of subject thickness, it is possible that the energy input to the x-ray tube in one frame may be decreased. The automatic control response was measured after rapid fluctuation in subject thickness. Two inverter-type x-ray generators with different automatic control units were studied. The older control unit changes exposure by tube voltage and tube current, while the newer one changes exposure dose by tube voltage, tube current and irradiation time. The maximum rate of change in tube voltage is greater with the newer control unit. In addition, the actual tube current response of the newer control unit in increasing nominal value is faster than the older one. In the new control unit, for each pulse, irradiation is cut off by means of a signal that the exposure has reached the proper value. Thus given the same differential in subject thickness, the newer control unit resumed stability faster than the older one.