Abstract
We investigated the removal of alimentary tract signal in T_2 echo planar imaging using an oral positive contrast agent for the alimentary tract mainly consisting of ferric ammonium citrate(FerriSeltz:Otsuka Pharmaceutical Co., Ltd.). It has been reported that the FerriSeltz preparation is useful as negative contrast agent because of its enhancing effect of reducing T_2 at high concentrations. However, it was shown to enhance susceptibility artifact in echo planar imaging(EPI). Thus, it is considered difficult to use FerriSeltz at high concentrations in EPI. In contrast, in IR EPI(TI 150 and 200 ms)high concentrations of FerriSeltz are not required, and it is possible to achieve excellent signal suppression at an ordinary concentration of 5.97 mmol/1. This might be due to the fact that the bounce point of FerriSeltz aqueous solution reached about 150 to 200 ms. At present, no effective negative contrast agent for alimentary tract MRI is available. Therefore, it is thought that IR EPI combined with FerriSeltz is a useful technique, since it can eliminate the signal of the alimentary tract at an ordinary concentration.