Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Moving-bed Infusion-tracking Three-dimensional MR Angiography by Applying Conventional CP Body Array Coil
MITSUJI HIGASHIDAMASARU YAMAZAKINOBUKO HANAOKAMASACHIKA SHOUGAKIKIMITO IMADA
Author information
JOURNAL FREE ACCESS

1999 Volume 55 Issue 12 Pages 1226-1231

Details
Abstract

We devised a new method of Mobi-track MRA by applying the conventional CP body array coil and our own sliding table. This method was applied for MR angiography of lower leg arteries combined with a subtraction technique. The performance of our method, which uses the CP body array coil, was examined in regard to misregistration of difference images, signal-to-noise ratio, contrast-to-noise ratio, and signal detectability, for comparison with that of the method using the body coil. Experimental results showed that our new method was superior to the method using the body coil. In conclusion, we consider our new method, which employs the CP body array coil and sliding table, to be very useful for three-dimensional MR angiography of lower leg arteries.

Content from these authors
© 1999 Japanese Society of Radiological Technology