Improvement of Image Quality in Chest MDCT Using Nonlinear Wavelet Shrinkage with Trimmed-thresholding
NARUOMI YASUDAYOKO ISHIKAWAYOSHIE KODERA
Author information
JOURNAL FREE ACCESS

2005 Volume 61 Issue 12 Pages 1599-1608

Details
Abstract
Multidetector-row computed tomography(MDCT)has dramatically increased the speed of scanning, and allows high-resolution imaging compared with conventional single detector-row CT(SDCT). However, the use MDCT makes use of an increase in volume scanning, and causes a simultaneous increase in radiation dose to the patient. Thus, the radiation dose from the X-ray CT has become a problem in recent years. In this study, nonlinear wavelet-based edge preservation de-noising using trimmed-thresholding was applied to reconstructed low-dose chest MDCT images, and optimal wavelet processing including wavelet functions and thresholding methods was examined. Moreover, the usefulness of the de-noising for reducing radiation dose was examined. As a result of optimized edge preservation de-noising, noise reduction was achieved with little deterioration in image quality, and the wavelet function used at that time was Coiflet's with shorter support. As a result, almost the same quality of reconstructed image of the chest phantom was obtained for conventional scanning and low-dose scanning with the wavelet de-noising method using trimmed-thresholding. That is, the radiation dose from MDCT could be reduced using this wavelet-based de-noising method.