Examination of Phantom Scatter Factors in Irregular Fields with Enhanced Dynamic Wedge
TOMOHARU SATOTOSHIHIRO ISHIDAMIWA SASAKITSUGIO OOTSUKA
Author information
JOURNAL FREE ACCESS

2005 Volume 61 Issue 2 Pages 253-259

Details
Abstract
Use of an enhanced dynamic wedge (EDW) changes the size of the irradiation field in real time because of the shift of the flexible aperture during irradiation. Therefore, when EDW is used, it is speculated that the scatter factors proposed by Khan et al. may differ from that obtained in an open irradiation field. The aim of the present study was to investigate whether the total scatter factors (S_<cp>) can also be separated into the collimator scatter factors (S_c) and phantom scatter factors (S_p) when using EDW, and whether S_p in an irregular irradiation field using EDW and one without the use of EDW can likewise be handled. Results indicated that the values of S_<cp> and S_c in a square or rectangular irradiation field using EDW tended to differ from those in an open irradiation field, whereas S_p was almost the same, with a slight variation of about 0.5%, indicating that S_<cp> could be separated into S_c and S_p. However, for an irregular irradiation field, the variations in S_p exceeded 1% in some cases irrespective of the use or non-use of EDW. Depending on the size and geometry of the irradiation field, it was within the range of the degree of uncertainty of 2.5% in the evaluation of output radiation dosage, which was described in AAPM Report 13. Therefore, the theory proposed by Khan et al. was found to be valid also for an irregular irradiation field using EDW, which could be handled in the same manner as in the irregular irradiation field without using EDW.
Content from these authors
© 2005 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top