Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Original
Reduction of Vascular Artifact on T1-weighted Images of the Brain by Using Three-dimensional Double IR Fast Spoiled Gradient Echo Recalled Acquisition in the Steady State (FSPGR) at 3.0 Tesla
Yasuhiro FujiwaraTsuyoshi MatsudaYoshiyuki IshimoriIsao YamaguchiHiroyuki HayashiYusuke OokoshiYuriko OotaniTosiaki MiyatiHirohiko Kimura
Author information
JOURNAL FREE ACCESS

2007 Volume 63 Issue 11 Pages 1239-1245

Details
Abstract

The purpose of this study was to decrease vascular artifacts caused by the in-flow effect in three-dimensional inversion recovery prepared fast spoiled gradient recalled acquisition in the steady state (3D IR FSPGR) at 3.0 Tesla. We developed 3D double IR FSPGR and investigated the signal characteristics of the new sequence. The 3D double IR FSPGR sequence uses two inversion pulses, the first for obtaining tissue contrast and the second for nulling vascular signal, which is applied at the time of the first IR period at the neck region. We have optimized scan parameters based on both phantom and in-vivo study. As a result, optimized parameters (1st TI=700 ms, 2nd TI=400 ms) successfully have produced much less vessel signal at reduction than conventional 3D IR FSPGR over a wide imaging range, while preserving the signal-to-noise ratio (SNR) and gray/white matter contrast. Moreover, the decreased artifact was also confirmed by visual inspection of the images obtained in vivo using those parameters. Thus, 3D double IR FSPGR was a useful sequence for the acquisition of T1-weighted images at 3.0 Tesla. (Article in Japanese)

Content from these authors
© 2007 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top