A Simulation Study of Radiographic Image Quality Measurement Based on Transmitted Information
Yongbum LeeDu-Yih TsaiEri Matsuyama
Author information
JOURNAL FREE ACCESS

2007 Volume 63 Issue 3 Pages 341-344

Details
Abstract
Although radiographic image quality is considered difficult to evaluate in a straightforward and systematic manner, it may be possible by using an index of transmitted information. As a preliminary study, relations between transmitted information and two image characteristics, namely, image noise and image blurring, were evaluated by simulation. The value of transmitted information was decreased if image noise and image blurring increased. The relationships were corroborated on an experimental basis. This paper suggests the possibility of a simple, straightforward method for synthetically evaluating radiographic images by a single parameter in terms of transmitted information.