Bedside Chest Radiography and Optimization of Radiographic Conditions in the NICU with the Fuji Computed Radiography System
Shoji TaniChiharu TanabeShuji AbeKazuhiro Mihara
Author information
JOURNAL FREE ACCESS

2008 Volume 64 Issue 1 Pages 15-24

Details
Abstract
We devised an application that uses the “Maximum value reading method (AutoIV)” for bedside chest radiography in a neonatal intensive care unit (NICU) that used the Fuji Computed Radiography (FCR) System. The application, named AutoIV-N, uses the relationship for density correction between radiographic conditions (mAs) and the Display Parameter (GS). GS=f (mAs) can be considered the relationship that connects FCR and the X-ray generator. When AutoIV-N is used, radiographic image contrast does not change. Further, radiographic image density fluctuation can be eliminated by random elements such as X-ray output fluctuation of the X-ray generator and the decline of photo-stimulated luminescence caused by fading of the imaging plate. Accordingly, image recording that is suitable for follow-up chest radiography is made possible. We choose nine patients and performed a comparison of radiographic density fluctuation in AutoIV-N and Fix. AutoIV-N was found to be more stable than Fix. It is possible to use the radiographic imaging condition that is optimized for all patients in the NICU by AutoIV-N. This facilitates radiation exposure optimization in medicine. (Article in Japanese)