Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Clinical Technology
Application of the Grayscale Standard Display Function to General Purpose Liquid-crystal Display Monitors for Clinical Use
Nobukazu TanakaKentaro NakaMasaki SueokaYoshiharu HigashidaJunji Morishita
Author information
JOURNAL FREE ACCESS

2010 Volume 66 Issue 1 Pages 25-32

Details
Abstract

Interpretations of medical images have been shifting to soft-copy readings with liquid-crystal display (LCD) monitors. The display function of the medical-grade LCD monitor for soft-copy readings is recommended to calibrate the grayscale standard display function (GSDF) in accordance with the guidelines of Japan and other countries. In this study, the luminance and display function of five models of eight general purpose LCD monitors were measured to gain an understanding of their characteristics. Moreover, the display function (gamma 2.2 or gamma 1.8) of general purpose LCD monitors was converted to GSDF through the use of a look-up table, and the detectability of a simulated lung nodule in the chest x-ray image was examined. As a result, the maximum luminance, contrast ratio, and luminance uniformity of general purpose LCD monitors, except for one model of two LCD monitors, met the management grade 1 standard in the guideline JESRA X-0093-2005. In addition, the detectability of simulated lung nodule in the mediastinal space was obviously improved by converting the display function of a general purpose LCD monitor into GSDF.

Content from these authors
© 2010 Japanese Society of Radiological Technology
Previous article Next article
feedback