Variation in the Quality of CT Images of the Upper Abdomen When CT Automatic Exposure Control Is Employed
Isao AizawaYoshihisa MuramatsuKeiichi NomuraFuminori Shimizu
Author information
JOURNAL FREE ACCESS

2010 Volume 66 Issue 3 Pages 185-192

Details
Abstract
The aim of this study was to analyze the reason for variation of image quality in the upper abdomen CT with the use of CT-AEC. The CT investigated was 3D modulation in the 16MDCT and LSCT phantom was used to simulate the patient. When there was a phase difference, an image noise increase of around 15% at the maximum was accepted. It is concluded that the major reason for variation in image quality is respiratory motion and the importance of respiration control must be recognized.