Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Originals
Examination of Motion Artifacts for Helical and Non–helical Scanning Modes in Head CT
Ichiro FujimuraKatsuhiro IchikawaShoichi TerakawaTakanori HaraYohei Miura
Author information
JOURNAL FREE ACCESS

2011 Volume 67 Issue 6 Pages 640-647

Details
Abstract

For head computed tomography (CT), non-helical scanning has been recommended even in the widely used multi-slice CT (MSCT). Also, an acute stroke imaging standardization group has recommended the non-helical mode in Japan. However, no detailed comparison has been reported for current MSCT with more than 16 slices. In this study, we compared the non-helical and helical modes for head CT, focusing on temporal resolution and motion artifacts. The temporal resolution was evaluated by using temporal sensitivity profiles (TSPs) measured using a temporal impulse method. In both modes, the TSPs and temporal modulation transfer factors (MTFs) were measured for various pitch factors using 64-slice CT (Aquilion 64, Toshiba). Two motion phantoms were scanned to evaluate motion artifacts, and then quantitative analyses for motion artifacts and helical artifacts were performed by measuring multiple regions of interest (ROIs) in the phantom images. In addition, the rates of artifact occurrence for retrospective clinical cases were compared. The temporal resolution increased as the pitch factor was increased. Remarkable streak artifacts appeared in the non-helical images of the motion phantom, in spite of the equivalent effective temporal resolution. In clinical analysis, results consistent with the phantom studies were shown. These results indicated that the low pitch helical mode would be effective for emergency head CT with patient movement.

Content from these authors
© 2011 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top