Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Originals
Positional Accuracy and Quality Assurance of Backup JAWs Required for Volumetric Modulated Arc Therapy
Daisaku TatsumiRyosei NakadaAkinori IenagaAkane YomodaMakoto InoueTakao IchidaMasako Hosono
Author information
JOURNAL FREE ACCESS

2012 Volume 68 Issue 11 Pages 1486-1491

Details
Abstract
The tolerance of the Backup diaphragm (Backup JAW) setting in Elekta linac was specified as 2 mm according to the AAPM TG-142 report. However, the tolerance and the quality assurance procedure for volumetric modulated arc therapy (VMAT) was not provided. This paper describes positional accuracy and quality assurance procedure of the Backup JAWs required for VMAT. It was found that a gap-width error of the Backup JAW by a sliding window test needed to be less than 1.5 mm for prostate VMAT delivery. It was also confirmed that the gap-widths had been maintained with an error of 0.2 mm during the past one year.
Content from these authors
© 2012 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top