Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Note
Evaluation of Imaging Quality and ADC for the Tetrahedral Gradients of Diffusion Weighted Imaging
Natsumi YamaguchiAkio OguraNorio HayashiDaisuke KoyamaOsakabe KoukiIsamu Hatano
Author information
JOURNAL FREE ACCESS

2015 Volume 71 Issue 7 Pages 595-604

Details
Abstract
The diffusion weighted images (DWIs) are used widely in clinical practice. Recently, the gradient overplus imaging was developed in addition to orthogonal technique for the duration method of motion-probing gradient pulses. The gradient overplus technique can reduce the load of a gradient strength, and enables setting of the shortest TE compared with the orthogonal gradients, and moreover it has the information on diffusion tensor. The apparent diffusion coefficient (ADC) values, signal-to-noise ratio (SNR), and geometric distortion were compared for the two duration methods of DWI using four different phantoms of coefficient of viscosity. In the gradient overplus, the time of smallest TE becamesmalle r than theorthogonal gradients according to theincre aseof theb value. As a result, SNR increased by shortening of TE, and thus the geometric distortion was improved. Moreover, the ADC value did not show any difference in the two duration methods of DWI. The gradient overplus technique is a more useful techniquein a clinical study than theorthogonal gradients becausethequality of imageis improved, and theADC valueis thesameas theorthogonal gradients.
Content from these authors
© 2015 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top