Evaluation of an Experimental Production Wireless Dose Monitoring System for Radiation Exposure Management of Medical Staff
Toshioh FujibuchiHiroo MurazakiTaku KuramotoYoshiyuki UmedzuYung Ishigaki
Author information
JOURNAL FREE ACCESS

2015 Volume 71 Issue 8 Pages 691-696

Details
Abstract
Because of the more advanced and more complex procedures in interventional radiology, longer treatment times have become necessary. Therefore, it is important to determine the exposure doses received by operators and patients. The aim of our study was to evaluate an experimental production wireless dose monitoring system for pulse radiation in diagnostic X-ray. The energy, dose rate, and pulse fluoroscopy dependence were evaluated as the basic characteristics of this system for diagnostic X-ray using a fully digital fluoroscopy system. The error of 1 cm dose equivalent rate was less than 15% from 35.1 keV to 43.2 keV with energy correction using metal filter. It was possible to accurately measure the dose rate dependence of this system, which was highly linear until 100 μSv/h. This system showed a constant response to the pulse fluoroscopy. This system will become useful wireless dosimeter for the individual exposure management by improving the high dose rate and the energy characteristics.