Estimation of Maximum Entrance Skin Dose during Cerebral Angiography
Satoru KawauchiTakashi MoritakeMikito HayakawaYusuke HamadaHideyuki SakumaShogo YodaMasayuki SatohLue SunYasuhiro KoguchiKeiichi AkahaneKoichi ChidaYuji Matsumaru
Author information
JOURNAL FREE ACCESS

2015 Volume 71 Issue 9 Pages 746-757

Details
Abstract
Using radio-photoluminescence glass dosimeter, we measured the entrance skin dose (ESD) in 46 cases and analyzed the correlations between maximum ESD and angiographic parameters [total fluoroscopic time (TFT); number of digital subtraction angiography (DSA) frames, air kerma at the interventional reference point (AK), and dose-area product (DAP)] to estimate the maximum ESD in real time. Mean (± standard deviation) maximum ESD, dose of the right lens, and dose of the left lens were 431.2±135.8 mGy, 33.6±15.5 mGy, and 58.5±35.0 mGy, respectively. Correlation coefficients (r) between maximum ESD and TFT, number of DSA frames, AK, and DAP were r=0.379 (P<0.01), r=0.702 (P<0.001), r=0.825 (P<0.001), and r=0.709 (P<0.001), respectively. AK was identified as the most useful parameter for real-time prediction of maximum ESD. This study should contribute to the development of new diagnostic reference levels in our country.