Multicenter Study on Evaluation of the Entrance Skin Dose by a Direct Measurement Method in Cardiac Interventional Procedures
Mamoru KatoKoichi ChidaTakashi MoritakeYasuhiro KoguchiYuji KagaHajime SakamotoAtsuko TsukamotoSatoru KawauchiKazuma MatsumotoMitsuaki MatsumuraHajime OosakaHideto Toyoshima
Author information
JOURNAL FREE ACCESS

2016 Volume 72 Issue 1 Pages 73-81

Details
Abstract
Deterministic effects have been reported in cardiac interventional procedures. To prevent radiation skin injuries in percutaneous coronary intervention (PCI), it is necessary to measure accurate patient entrance skin dose (ESD) and maximum skin absorbed dose (MSD). We measured the MSD on 62 patients in four facilities by using the Chest-RADIREC system. The correlation between MSD and fluoroscopic time, dose area product (DAP), and cumulative air kerma (AK) showed good results, with the correlation between MSD and AK being the strongest. The regression lines using MSD as an outcome value (y) and AK as predictor variables (x) was y=1.18x (R2=0.787). From the linear regression equation, MSD is estimated to be about 1.18 times that of AK in real time. The Japan diagnostic reference levels (DRLs) 2015 for IVR was established by the use of dose rates using acrylic plates (20- cm thick) at the interventional reference point. Preliminary reference levels proposed by International Atomic Energy Agency (IAEA) were provided using DAP. In this study, AK showed good correlation most of all. Hence we think that Japanese DRLs for IVR should reconsider by clinical patients' exposure dose such as AK.
Content from these authors
© 2016 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top