Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Originals
A Simple Method for Computationally Generating Metal Artifacts in CT Images for Treatment Planning: A Pilot Phantom Study
Daisuke MugishimaAkihiro NaritaMasaki Ohkubo
Author information
JOURNAL FREE ACCESS

2021 Volume 77 Issue 5 Pages 445-453

Details
Abstract

Purpose: In treatment planning for radiation therapy, the use of computed tomography (CT) images including metal artifacts causes a reduction in the dose calculation accuracy. In clinical practice, the artifacts are manually contoured and assigned an appropriate fixed CT number. To validate the procedure, images taken before and after metal insertion into a patient are required, which may be impractical. We propose a simple method for computationally generating metal artifacts in clinical images. Methods: In the proposed method, a clinical image free of metal artifacts is used. To simulate metal inside a patient, CT numbers of a region in the image are replaced with a fixed extremely high value. A sinogram is created by the forward projection of the image. Data values of the sinogram in the metal region are converted into smaller values. From the sinogram, an image including artifacts is reconstructed with the filtered back projection. Results: The simulated artifacts consisted of dark and bright bands and were observed to be similar to the actual metal artifacts. CT numbers in multiple small regions of interest in the image obtained by the proposed method showed a good agreement with those in the actual image. Conclusion: The proposed method was demonstrated to generate the metal artifacts additionally on the clinical images. The method would be potentially applicable to a validation study for the clinical procedure of manually contouring and assigning CT numbers to metal artifacts.

Content from these authors
© 2021 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top