2023 Volume 79 Issue 2 Pages 121-127
Purpose: This study deals with the conversion of the image quality figure (IQFinv) values with CDMAM Analyser ver. 1 (old analysis software) used to analyse the images of CDMAM 3.4 phantom into the IQFinv values with a new analysis software, that is, CDMAM Analyser (ver. 2 and ver. 2.3), which is a software improved to be closer to human visual evaluation. Methods: The IQFinv values were calculated using four mammography units and three analysis software. The IQFinv values using the old and new software were compared. Results: The IQFinv values had no difference according to mammography units and beam quality. The conversion formula of IQFinv from CDMAM Analyser ver. 1 to CDMAM Analyser ver. 2 was derived. Furthermore, the conversion formula of IQFinv from CDMAM Analyser ver. 1 to CDMAM Analysis ver. 2.3 was also derived. Conclusion: Using each conversion formula, the IQFinv values using the old software can be converted to those using the new software. There was a slight difference between the IQFinv values using the new software.