Impact of Different Acquisition Modes on Image Quality and Quantitative Accuracy for Bone SPECT Using Ordered Subset Conjugate-gradient Minimizer
Yui Orisaka Takayuki ShibutaniTakahiro KonishiAnji YokouchiKatsutoshi FujiwaraHiroto Yoneyama
Author information
JOURNAL FREE ACCESS

2024 Volume 80 Issue 11 Pages 1191-1197

Details
Abstract

We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and 99mTc was filled with the SIM2 Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuousview, continuoustotal) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuousview, and continuoustotal methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.

Content from these authors
© 2024 Japanese Society of Radiological Technology
Previous article Next article
feedback
Top