THE JOURNAL OF JAPAN SOCIETY FOR CLINICAL ANESTHESIA
Online ISSN : 1349-9149
Print ISSN : 0285-4945
ISSN-L : 0285-4945
Other Reports
The Inhibitory Effect of Fluorescein on the Measurements of Hemoglobin Concentration by CO-oximetry
Yoko SAKAMOTOYoshinori MAEDAHiroko ONOKeiko MIYAZAKIShozaburo JOTAKIYoshikado MIYAGAWA
Author information
JOURNAL FREE ACCESS

2017 Volume 37 Issue 2 Pages 189-192

Details
Abstract

Fluorescein, a fluorescent dye, has recently begun to be used in craniotomy to improve the tumor resection rate and avoid neurological complications.

Errors of CO-oximetry occurred in 3 patients after fluorescein administration.

Three patients were administered fluorescein at 1,000 mg intravenously. After administration, measurement of CO-oximetry using RapidPoint 405TM(SIEMENS)including the hemoglobin concentration became impossible.

The errors are believed to have occurred as the measurement wavelength of CO-oximetry was near the wavelength of the absorption peak of fluorescein.

Content from these authors
© 2017 by The Japan Society for Clinical Anesthesia
Previous article Next article
feedback