It was speculated that waste sampling gas from CO2 monitor might be another source of air pollution in the operating room.
We measured N2O concentrations in the operating room when CO2 monitor was in use.
In the poorly ventilated room, N2O concentrations reached near 100ppm due to waste gas from CO2 monitor.
Thus, the waste gas from CO2 monitor is another source of air pollution in the operating room, and waste gas should be scavenged.