Journal of the Japan Society for Precision Engineering
Online ISSN : 1882-675X
Print ISSN : 0912-0289
ISSN-L : 0912-0289
Measurement of Straight Motion Accuracy Using the Improved Sequential Three-Point Method
Joji YAMAGUCHI
Author information
JOURNAL FREE ACCESS

1993 Volume 59 Issue 5 Pages 773-778

Details
Abstract
The purpose of the paper is to realize precise measurement of straight motion accuracy over long travel. The reversal method as well as the sequential three-point method is conventional. However, it is difficult to measure motion accuracy over long travel precisely using either method. The paper proposes a new precise measuring method which will be called the improved sequential three-point method. In this method, the principles of the reversal method and the sequential three-point method are introduced. Straight motion accuracy of a slide table was measured over 850mm travel using the improved sequential three-point method. The standard deviations from 12times measurements were 1.0μm in horizontal position and 2.0arc sec in yaw angle.
Content from these authors
© by The Japan Society for Precision Engineering
Previous article Next article
feedback
Top