The Journal of the Japan Society for Respiratory Endoscopy
Online ISSN : 2186-0149
Print ISSN : 0287-2137
ISSN-L : 0287-2137
Clinical Evaluation of Percutaneous Needle Aspiration Cytology by Gyro-Fluoroscope
Shinya KusachiMasumi KurashigeHideaki SuzukiYasumasa TaguchiMinoru KuritaMasatake SuzukiKunihiko KawaiYoshinobu Sumiyama
Author information
JOURNAL FREE ACCESS

1988 Volume 10 Issue 2 Pages 148-152

Details
Abstract

The results of 21 percutaneous needle aspiration cytology procedure in case of peripheral lung lesions using a gyro-fluoroscope were analyzed. Of 19 malignant lung lesions, 18 were diagnosed cytologically, and two tuberculomas were diagnosed bacteriologically. Of 21 patients, in 20 (95.2%) were diagnosed correctly. Slight transient hemoptysis was experienced in only one case, but there was no case of pneumothorax or dissemination. The gyro-fluoroscope yielded high accuracy with few complications.

Content from these authors
© 1988 The Japan Society for Respiratory Endoscopy
Previous article Next article
feedback