Research Note
Domestic Comparison of Relative Spectral Responsivity Measurements for Illuminance Meters
Kazuaki OHKUBOMotomi HORIUCHIYasuo NAKAGAWAHitoshi TOZAWAKenji KOBAYASHIIzumi HORlENaomich’ CHIDA
Author information
JOURNAL FREE ACCESS

2000 Volume 24 Issue 1 Pages 1_66-1_72

Details
Abstract
A comparison of relative spectral responsivity measurements of illuminance meters was carried out among five companies — TOPCON Corp., HIOKI E.E. Corp., Minolta Co., Ltd., Yokogawa M&C Corp., and Matsushita Electric Industrial Co., Ltd. The spectral responsivitys of the illuminance meters were measured by each company's spectral responsivity measurement system calibrated to each company's standard. As a result, the measured values of fs (fs is defined as the relative spectral responsivity error in Japanese Industrial Standard JIS-C-1609) differed within 1.7. The fs is identical with the f1' recommended by the CIE.