Journal of Microorganism Control
Online ISSN : 2758-6391
Print ISSN : 2758-6383
Note
External quality control survey on identification of microorganisms using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
KAZUYUKI SOGAWAAZUMI FUJINAGAHAJIME OKUMURAMAKIKO KIYOSUKESYOTA MURATAKOJI KUSABAKOHEI UECHIKAZUKI HORIUCHIKAZUNARI YASUDAMASAMI MURAKAMITOMOHIRO NAKAYAMA
Author information
JOURNAL FREE ACCESS

2024 Volume 29 Issue 1 Pages 49-53

Details
Abstract

Abstract: Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry( MALDI-TOF MS) is a bacterial typing tool that was approved as a medical device in 2011. However, external accuracy control examination of bacterial typing using mass spectrometry is still only performed on a small scale. In this study, E. faecium and S. maltophilia were selected and tested according to established procedures using Score Values at 228 institutions. The Score Values for MALDI Biotyper were 2.43±0.08 for E. faecium and 2.38±0.08 for S. maltophilia; and those for VITEK MS/PRIME were 99.9±0.0 for E. faecium and S. maltophilia. These results suggest that it is useful to evaluate external accuracy control with Score Values using the procedures we have developed.

Content from these authors
© 2024 The Society for Antibacterial and Antifungal Agents, JAPAN
Previous article
feedback