Noninvasive continuous blood pressure monitoring by the ClearSight system during robot‐assisted laparoscopic radical prostatectomy
Yoko SakaiTsutsumi Yasuo M.Takuro OyamaChiaki MurakamiNami KakutaKatsuya Tanaka
Author information
JOURNAL FREE ACCESS

2018 Volume 65 Issue 1.2 Pages 69-73

Details
Abstract

Robot‐assisted laparoscopic radical prostatectomy (RALRP) is commonly performed in the surgical treatment of prostate cancer. However, the steep Trendelenburg position (25) and pneumoperitoneum required for this procedure can sometimes cause hemodynamic changes. Although blood pressure is traditionally monitored invasively during RALRP, the ClearSight system (BMEYE, Amsterdam, The Netherlands) enables a totally noninvasive and simple continuous blood pressure and cardiac output monitoring based on finger arterial pressure pulse contour analysis. We therefore investigated whether noninvasive continuous arterial blood pressure measurements using the ClearSight system were comparable to those obtained invasively in patients undergoing RALRP. Ten patients scheduled for RALRP with American Society of Anesthesiologists physical status I‐II were included in this study. At each of the seven defined time points, noninvasive and invasive blood pressure measurements were documented and compared in each patient using Bland‐Altman analysis. Although the blood pressure measured with the ClearSight system correlated with that measured invasively, a large difference between the values obtained by the two devices was noted. The ClearSight system was unable to detect blood pressure accurately during RALRP, suggesting that blood pressure monitoring using this device alone is not feasible in this small patient population. J. Med. Invest. 65:69‐73, February, 2018

Content from these authors
© 2018 by The University of Tokushima Faculty of Medicine
Previous article Next article
feedback
Top