Effective Visualization of the Intracranially Implanted Neurosurgical Devices with Photon-counting Detector Computed Tomography:A Technical Note
Tadasuke Tominaga, Takahiro Yokoyama, Suguru Nakamura, Hanako Morisawa, Tatsuya Hoshino, Takakazu Kawamata, Naohisa Miura
Author information
JOURNAL FREE ACCESS

2025 Volume 72 Issue 3.4 Pages 463-466

Details
Abstract

Photon-counting detector computed tomography (PCD-CT) allows for attainment of images with excellent spatial resolution at reduced radiation doses. However, practical utility of this diagnostic modality in neurosurgery is still not defined well. Herein, the authors report their initial experience with the use of postoperative PCD-CT (helical pitch, 0.85;gantry rotation time, 0.5 s;tube voltage, 120 kV) in a single-acquisition high-resolution mode with the iterative metal artifact reduction and attainment of the reconstructed images with slice thickness of 0.2 mm for assessment of the shunt valve pressure adjustment system and endovascular flow-diverting stent. In both cases, intracranially implanted neurosurgical devices were effectively visualized. J. Med. Invest. 72 : 463-466, August, 2025