Journal of Microwave Surgery
Online ISSN : 1882-210X
Print ISSN : 0917-7728
ISSN-L : 0917-7728
Ariticles
An experimental study about improvement of operation and safety of the microwave coagulator
Suminori KitadaKoichi WakikaidoMasaki YamadaEiji ShimadaKatsuyoshi Tabuse
Author information
JOURNAL FREE ACCESS

1996 Volume 14 Pages 33-36

Details
Abstract
The level of reflected wave that is appeared when microwave was output is different by the condition of junction which is connected the output cable or electrode. We devised the equipment to prevent miss-operation by using this characteristic. As a result, we made it possible to give warning for miss-operation by using the wave which is subtracted reflected wave from forward wave. And we tried to find the timing that coagulation finishes, using the characteristic that level of reflected wave changes with spread the coagulated layer.