Journal of Microwave Surgery
Online ISSN : 1882-210X
Print ISSN : 0917-7728
ISSN-L : 0917-7728
Electromagnetic Interference in Cardiac pacemakers
Electrocautery and Microwave scalpel
Seiki YamamotoKatsuyoshi TabuseMasahiro SakaguchiNakahiro ShimotsumaKouji MinamiMiho NagahamaKazuo AriiMasami Oka
Author information
JOURNAL FREE ACCESS

1992 Volume 10 Pages 59-66

Details
Abstract

In Japan there are more than 15000 patients with cardiac pacemaker in a year. General surgical procedures in pacemaker patients have become common in recent years. General surgeon must have enough knowledge about surgical problem in pacemaker patients. It has been one of the important problem that the use of electrocautery sometimes interferes in the heart rhythm by pacemaker although there is no report, whether microwave scalpel has influence on the pacemaker rhythm or not. Recently we have alternately used electrocautery and new microwave scalpel (Blade type) on same patient during gastrectomy. In this paper, The influence of the electrocautery and microwave scalpel on the pacemaker function is reported respectively. It is demonstrated that a microwave scalpel has not influenced on the pacemaker rhythm as compared with electrocautery.