A Fast Response UV-Absorption Ozone Photometer using a Differentially-Phased Triple-Cell System
Yukitomo Tsutsumi
Author information
JOURNAL FREE ACCESS

1997 Volume 75 Issue 5 Pages 969-974

Details
Abstract
A new fast-response ozone photometer which uses ultraviolet (UV)-absorption is described. Its three UV-absorption cells each cycle among three modes: reference, sample and transition modes. The three cells are differentially phased, enabling the ozone concentration to be determined in a cell in one second without interruption. The system is efrective for measurements requiring a fast response, such as measurements on aircraft. We made a trial instrument that adopts this system and confirmed its theoretically predicted operation. The trial instrument showed a significantly faster response than earlier types of ozone photometers. This system may be applicable to other gas measurements that uses UV-absorption.
Content from these authors