Journal of the Magnetics Society of Japan
Online ISSN : 1880-4004
Print ISSN : 0285-0192
ISSN-L : 0285-0192
Magneto-Optics
A New Measurement System of the Surface Magneto-Optic Kerr Effect (SMOKE)
Y. SuzukiM. OhhirakiH. KikuchiT. KatayamaN. Koshizuka
Author information
JOURNAL OPEN ACCESS

1989 Volume 13 Issue 2 Pages 167-170

Details
Abstract
A new measurement system of surface magnetooptic Kerr effect (SMOKE) has been developed. Combination of a polarization modulating method using electric detection and signal processing using a computer system provides very high sensitivity, 10-4 degrees, and a good signal to noise ratio. Moreover, electrical calibration using a Faraday cell makes it possible to measure the absolute value of a Kerr rotation angle of the order of 10-4 degrees. SMOKE hysteresis of 2-atomic layer of fcc-Co grown on a Cu(100) surface is successfully measured and its Kerr rotation angle is about 5×10-4 degrees.
Content from these authors
© 1989 by The Magnetics Society of Japan
Previous article Next article
feedback
Top