Journal of Neuroendovascular Therapy
Online ISSN : 2186-2494
Print ISSN : 1882-4072
ISSN-L : 1882-4072
Original Articles
Analysis of the Anatomical Factors Affecting Ability to Navigate Penumbra Catheter through Internal Carotid Siphon
Emi KuriyamaYoshikazu MatsudaTakumi KawaguchiRie YakoNaoyuki Nakao
Author information
JOURNAL OPEN ACCESS

2020 Volume 14 Issue 5 Pages 169-176

Details
Abstract

Objective: The efficacy and safety of aspiration thrombectomy using Penumbra catheter with acute large vessel occlusion in the anterior circulation have been reported in previous studies. In some cases, the carotid siphon (CS) is elongated, and with this anatomy, especially where there is bifurcation of the ophthalmic artery (OA), navigation of Penumbra catheters into distal internal carotid artery (ICA) is interrupted, which is known as the ‘ledge effect’. We investigate the anatomical characteristics of CS that cause interruption of navigation of the Penumbra catheter.

Methods: Between January 2015 and March 2018, mechanical thrombectomy using Penumbra 60 was performed on 51 patients with middle cerebral artery (MCA) or intracranial ICA occlusion. Patients were divided into two groups: The ‘ledge-effect’ group those in whom the Penumbra catheter was unable to be navigated into the distal ICA through the CS, and ‘no ledge-effect’ group those in whom this was possible. The anatomical characteristics of CS, the diameter of ICA, diameter of OA, OA/ICA ratio and radius of the CS were evaluated using angiographical imaging.

Results: The ‘ledge-effect’ group numbered eight cases (17%). Only the value of the CS radius was significantly smaller in the ledge-effect group (p = 0.0019), other parameters were not significantly different between the groups. The cutoff radius value was 3.62 mm.

Conclusion: The most notable anatomical factor affecting possibility of navigation of the Penumbra catheter through the CS was the CS radius. This could be useful information when devices used in mechanical thrombectomy are selected.

Content from these authors
© 2020 The Japanese Society for Neuroendovascular Therapy

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top