Vessel Rupture Complication due to Ballooning of the Distal Segment of a JET 7 Xtra Flex Reperfusion Catheter during Angiogram
Takanori SanoKoichi HakozakiKazuto KobayashiTomonori IchikawaHiroshi TanemuraTomoki IshigakiFumitaka Miya
Author information
JOURNAL OPEN ACCESS

2022 Volume 16 Issue 4 Pages 218-224

Details
Abstract

Objective: We present a case of intraprocedural device malfunction related to the JET 7 Xtra Flex reperfusion catheter during mechanical thrombectomy.

Case Presentation: A 92-year-old man presented with sudden right hemiparesis with a National Institutes of Health Stroke Scale score of 22. His left middle cerebral artery (M1) was occluded, and emergency mechanical thrombectomy was performed. After partial recanalization was achieved, angiography through a JET 7 Xtra Flex was attempted. After manual injection of contrast media via a 10-mL syringe through the JET 7 Xtra Flex, the catheter moved, jumping forward, and the distal tip of the catheter expanded and ruptured. This resulted in intracranial vessel damage and subsequent patient death.

Conclusion: Contrast media must not be injected through the JET 7 Xtra Flex. If contrast media needs to be injected for angiography during mechanical thrombectomy with a reperfusion catheter, it should always be through the guide catheter.

Content from these authors
© 2022 The Japanese Society for Neuroendovascular Therapy

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top