Journal of Neuroendovascular Therapy
Online ISSN : 2186-2494
Print ISSN : 1882-4072
ISSN-L : 1882-4072
Case Reports
Difficulties in withdrawing a stent delivery system using Mo. Ma Ultra: case report
Hidemitsu ADACHINobuyuki SAKAIHirotoshi IMAMURAShoichi TANIChiaki SAKAITatsuya ISHIKAWAYohei MINEHARUKatsunori ASAIHiroyuki IKEDATaku INADATakenori OGURATeishiki SHIBATAMikiya BEPPUYuji AGAWAKanpei SHIMIZUKenichi TODOShiro YAMAMOTO
Author information
JOURNAL OPEN ACCESS

2013 Volume 7 Issue 5 Pages 338-344

Details
Abstract
Objective: Mo. Ma Ultra (Mo. Ma) is used during carotid artery stenting (CAS) and, in two cases, we experienced difficulties withdrawing the stent delivery system. Here we report the mechanism and procedures that should be performed for safe withdrawal of the system.
Case presentation: Representative case 1: An 81-year-old male patient with symptomatic carotid stenosis who underwent CAS using Mo. Ma. During the operation, we encountered issues with withdrawing the stent delivery system.
Conclusion: The cause of difficulties in withdrawing the stent delivery system in CAS using Mo. Ma is the stent tip getting caught in the outlet port of Mo. Ma. The stent system can be safely withdrawn by resheathing the outer sheath of a Precise stent accurately.
Content from these authors
© 2013 The Japanese Society for Neuroendovascular Therapy

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback