Journal of Neuroendovascular Therapy
Online ISSN : 2186-2494
Print ISSN : 1882-4072
ISSN-L : 1882-4072

This article has now been updated. Please use the final version.

Usefulness of Preoperative Simulation Using a Stereolithographic 3D Printer in Cerebral Aneurysm Coil Embolization
Toshiyuki OndaTadashi NonakaTatsufumi NomuraShigeru InamuraOsamu HondaMasahiko Daibou
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: tn.2020-0151

Details
Abstract

Objective: We present a preoperative simulation of cerebral aneurysm coil embolization using a hollow model of cerebral blood vessels created by a stereolithography (SLA) 3D printer.

Case Presentation: The patient was a 66-year-old woman. During follow-up, coil embolization was planned for an expanding paraclinoid aneurysm. A hollow cerebral vascular model was created preoperatively using an SLA 3D printer. The catheter was malleable and inserted into the hollow model, which enabled the surgeons to confirm its movement, stability, and ease of insertion. In the surgical procedure, the catheter was easily inserted into the aneurysm without reshaping. The procedure was completed without stability problems.

Conclusion: The use of a hollow model of cerebral blood vessels was useful as a preoperative simulation and improved the safety of the procedure.

Content from these authors
© 2021 The Japanese Society for Neuroendovascular Therapy

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
feedback
Top