Journal of Nuclear Science and Technology
Online ISSN : 1881-1248
Print ISSN : 0022-3131
Automatic Measuring Instrument for Radon Daughters Concentration of Precipitation
Tsuguo NISHIKAWAMasayoshi AOKIShigeru OKABE
Author information
JOURNAL FREE ACCESS

1986 Volume 23 Issue 11 Pages 1001-1007

Details
Abstract

An automatic measuring instrument for the radon daughters concentration of precipitation has been constructed and used successfully. The precipitation collected by a polyethylene funnel during 15 min is led to a glass cell set in a well-type NaI (T1) scintillator, and after 5 min counting the water sample is pumped out. A sequence of operations of sampling, measurement, recording, draining out and correction for the volume of the water sample is performed automatically, and is repeated at every 15 min. The instrument has sensitivity and durability enough for practical use at outdoors. This instrument enables the measurement of radon daughters concentration as low as 0.5 Bq/ml at precipitation rate as small as 0.5 mm/h.
For the improvement of accuracy of the monitoring system of environmental γ-radiation, the effect of radon daughters accumulated on the ground surface by precipitation can be eliminated on real time by means of providing it with the output of this instrument.
This instrument can be used for the measurement not only of the radon daughters but also of the other radioactive nuclides in precipitation.

Content from these authors

This article cannot obtain the latest cited-by information.

© the Atomic Energy Society of Japan
Previous article Next article
feedback
Top