Moving Strip Technique of Electoron Beam Therapy
Kisho MATSUSHIMAHiroyuki WAKASANobuhiro OGURITakuichi KITAYAMAYoshitada NAKAGIRIYasutaka MIKAMIKeiji HASHIMOTOYoshio HIRAKIKaname AONO
Author information
JOURNAL FREE ACCESS

1984 Volume 96 Issue 11-12 Pages 1175-1182

Details
Abstract

The fieldsize in electron beam therapy is determined by the cone size. In cases of skin metastasis of a malignant tumor and so on, which need a large field size and whose area is much larger than the size of the cone, a large field size is usually produced by dividing the portals. However, the dose distribution at the border of the field becomes unequal, and hot and cold dose areas are produced according to the distance between portals. We tried the strip field technique in a large field along the long axis of the body in order to flatten the dose of the border employing the moving strip used for whole abdominal irradiation in ovarian cancer.
We set the film in Mix-DP and used the strip field technique with 2.5cm steps. We discussed the relationship between the interval (distance between portals) and the flattening of the dose within the field. Skin movement due to breathing and influences on the flattening of the dose were considered.
The proper flatness was obtained at depths of 0, 1, 2, and 3cm by setting the interval at 0.5cm. When skin movement was produced by breathing in ±1.5mm, the proper flaness was obtained also at a 0.5-cm interval. It seems that smoothing is increased by breathing.
An “electron beam moving strip” with a 2.5-cm step and 0.5-cm interval was clinically effective in the treatment of patients with skin metastasis of colon cancer.