Narrow Bore Packed Column Designed for Capillary Gas Chromatograph
Ikuo UETAYoshihiro SAITO
Author information
JOURNAL FREE ACCESS

2014 Volume 35 Issue 1 Pages 41-48

Details
Abstract

A narrow-bore packed column (packed-capillary column) that designed for a direct installation in conventional capillary gas chromatographic (GC) system had been developed. In this column, particulate sorbents that employed in typical GC packed columns were packed into a stainless-steel capillary of 1.0 mm i.d. By attaching a pair of stainless-steel capillaries of 1.27 mm o.d. at both ends of the column, the packed-capillary column can be directly installed into the conventional capillary GC system. In this article, fundamental of the packed-capillary column developed by our research group is mainly reviewed, along with the various applications, such as column switching analysis and evaluation of catalytic property based on the determinations of CO and CO2.

Content from these authors
© 2014 The Society for Chromatographic Sciences