Evaluation of Simultaneous Binocular Autorefractometry in Normal and Ametropic or Anisometropic Amblyopia
Takako TachikawaIkuko GotoMasao MatsubaraKazuno NegishiKenji OhnoToru NodaTakefumi HayashiKoji NishioYasufumi Fukuma
Author information
JOURNAL FREE ACCESS

2008 Volume 29 Issue 1 Pages 26-31

Details
Abstract

We measured refraction using a newly developed simultaneous binocular autorefraction device (BV-1000, Topcon) and compared the measurements with those obtained using a conventional monocular autorefractometer (KR-8100A, Topcon). Children with ametropic amblyopia without strabismus (15 patients, 30 eyes, 5.6±1.6 years old), anisometropic amblyopia (10 patients, 20 eyes, 7.6±3.3 years old), without amblyopia (20 patients, 40 eyes, 7.4±3.6 years old) and adults (10 patients, 20 eyes, 34.6±7.8 years old) were evaluated. Objective refraction before and after instillation of 1% cyclopentolate was measured using the two devices. In each group with and without amblyopia, the difference between pre- and post-cycloplegic refractive values was significantly smaller in BV-1000 measurement than in KR-8100A measurement. In anisometropic amblyopia, regarding the disparity between fellow eyes, the differences between pre- and post-cycloplegic measurements varied less in BV-1000 measurement than in KR-8100A measurement. Use of the BV-1000, enables accurate measurement of anisometropia without cycloplegia.

Content from these authors
© 2008 The Japanese Society of Ophthalmological Optics
Previous article Next article