Journal of Plastic and Reconstructive Surgery
Online ISSN : 2436-259X

This article has now been updated. Please use the final version.

Suitability of the Microvascular Anastomotic Coupler System for Small Spliced Vein Grafts
Yutaro YamashitaShinji NagasakaKazuhide MinedaYoshiro AbeIchiro Hashimoto
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: 2022-0011

Details
Abstract

Microvascular anastomotic coupler system is used in vascular surgical procedures. By using this device, vascular anastomosis can be performed faster and more easily. It is often used for venous anastomosis during microsurgical reconstruction. We performed distal bypass surgery on two patients with critical limb ischemia with spliced vein grafts using the microvascular anastomotic coupler system. The blood vessel diameters were 2 and 2.5 mm, respectively. Both grafts showed obstruction at 3 and 1 months, respectively, after surgery. The microvascular anastomotic coupler system may not be suitable for small-diameter spliced vein grafts.

Content from these authors
© 2022 Japan Society of Plastic and Reconstructive Surgery

JPRS is an Open Access journal distributed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. Anyone may download, reuse, copy, reprint, or distribute articles published in the Journal for non-profit purposes if they cite the original authors and source properly. If anyone remixes, transforms, or builds upon the material, you may not distribute the modified material.